Development and Validation of RP-HPLC method for the Estimation of Ritonavir in API and tablet Formulation
Avinash Birajdar*, Varsha Tegeli, Suyash Ingale, Gajanand Nangare
D.S.T.S. Mandal痴 College of Pharmacy, Solapur-413004, Maharashtra, India.
*Corresponding Author E-mail: avibirajdar1998@gmail.com
ABSTRACT:
A Reverse Phase High Performance Liquid Chromatographic method has been developed and validated for estimation of Ritonavir in API and Tablet formulation. The chromatographic separation was carried out using ZOROBAX Bonus-RP C-18 column (250 x 4.6mm, 5オm) as stationary phase and Methanol: Acetonitrile: 0.1% Trifluoroacetic acid water (81:9:10) as mobile phase at 1.0 ml/min flow rate and detection was carried out at 250 nm. The method was validated accordance to the Guidelines of International Council for Harmonization (ICH). Ritonavir have linearity in the concentration range of 50-150オg/ml with correlation coefficient (r2=1) respectively. Ritonavir eluted at 3.05 min respectively. The method is sensitive, precise and accurate. So, the method can be successfully applied for the routine analysis of Ritonavir in pharmaceutical formulations.
KEYWORDS: Ritonavir, RP-HPLC, Method development, Validation, 250nm and Diode Array Detector.
1.0. INTRODUCTION:
The chemical name of Ritonavir is (5S, 8S, 10S, 11S) -10- hydroxy- 2-methyl- 5- (1- methylethyl) - 1- [2- (1-methylethyl) - 4- thiazolyl] -3, 6-dioxo- 8, 11- bis (phenylmethyl)-2, 4, 7, 12- etraazatridecan- 13-oic acid 5-thiazolyl methyl ester. It is official in Indian Pharmacopoeia1 and United States Pharmacopoeia.2 It has the Molecular Formula is C37H48N6O5S2, Molecular Weight is 720.948 g/mol and structural formula shown in (Fig. 1).
Fig. 1: Structure of Ritonavir
Ritonavir3 is an antiretroviral drug from the protease inhibitor class used to treat HIV infection and AIDS. Ritonavir is frequently prescribed with Highly Active Anti-Retroviral Therapy, not for its antiretroviral action, but as it inhibits the same host enzyme that metabolizes other protease inhibitors. This inhibition leads to higher plasma concentrations of these latter drugs, allowing the clinician to lower their dose and frequency and improving their clinical efficacy. From the literature survey, it was found that Ritonavir estimated by analytical methods such as RP-HPLC method4, 5, 6, 7, 8, 9, 10,16, UV11-14 Spectroscopic LC-MS15 and HPTLC method.17 The developed method was simple, precise, specific and accurate. The statistical analysis proved that method is reproducible and selective for the analysis of Ritonavir in API and tablet formulations.
2.0 MATERIAL AND METHOD:
2.1 Chemicals and reagents
The drug Ritonavir was obtained as gift sample from Aadhar Life Sciences. HPLC grade Methanol, Acetonitrile, 0.1% Trifluoroacetic acid water Merck Specialities Pvt. Ltd., Mumbai, India.
2.2 Instruments:
Analytical balance (Aczet CY224C), HPLC (Agilent 1260 Infinity II autosampler), Diode Array Detector (G4212B), UV-VIS Double Beam Spectrophotometer 2201, Vortex machine (Remi CM 101 plus), Sonicator (Labman).
2.3 Chromatographic equipment and conditions:
Chromatographic analysis was performed on a Zorbax Bonus-RP (250 x 4.6 mm, 5オ). The mobile phase consists Methanol: ACN: 0.1% TFA Water (81: 9: 10) and was set at a flow rate of 1.0 ml/min. The mobile phase was ultrasonicated for degassing and filtered through 0.45オm membrane nylon filter using vacuum pump before pumping into HPLC system. The effluent was monitored by PDA detector at 250nm.
2.4 Preparation of solutions:
2.4.1Preparation of standard stock solution
10mg of Ritonavir API was transferred into 10 ml of volumetric flask containing 5 ml of diluents. The solution was ultrasonicated for 15 minutes. The volume was made up to the mark with diluents and vortex for 5 min. (Conc. = 1000 オg/ml). From that solution pipette out 1.0ml of in 10ml volumetric flask. Add 5 ml diluent and vortex; make up the volume with diluent. (Conc. = 100オg/ml)
2.4.2 Preparation of Drug Product stock solution
10 tablets were weighed and average tablet weight was calculated. These tablets were powdered in mortar and pestle. Powder equivalent to 10 mg Ritonavir as lable claim, was weighed and dissolved in 10ml Diluent. (Sample Conc. = 1000オg/ml) From that solution pipette out 1.0ml of in 10ml volumetric flask. Add 5 ml diluent and vortex; make up the volume with diluent. (Conc. = 100オg/ml)
2.4.3 Preparation of Mobile Phase
HPLC grade Methanol, Acetonitrile and 0.1% Trifluoroacetic acid water used for the preparation of mobile phase in the 81: 9: 10 ratio.
Preparation of 0.1% TFA Water: Accurately weigh 0.1ml of Trifluoroacetic acid in 1000ml of beaker and dissolved in 1000 ml Water. Filter twice using 0.45オ membrane filter and degas for 15 min. Each component of mobile phase filtered through 0.45オ membrane filter twice. Sonication is performed for 15mins.to degas the mobile phase
3.0 Validation of Analytical Method:
I. Specificity
The analytical parameter specificity is the ability to detect the analyte in presence of other components example impurities, degradants, matrix etc.
To provide an exact result which allows an accurate statement on the content or potency of the analyte in a sample. The specificity of the method was determined by comparing test results obtained from analysis of sample solution containing excipients with that of test results those obtained from standard drug and assay was also calculated. This solution was assayed at 250nm.
��������������������������������������� Sample area
Formula: % Assay = 末末末末末末末末末末末堀 100
��������������������������������������� Standard area
II. Linearity
Linearity of an analytical procedure is its ability to obtain test results which is directly proportional to the concentration (amount) of analyte present in the sample.
50, 75, 100, 125オg/ml solutions were prepared from the standard stock solution of Ritonavir (1000オg/ml). The solution were filtered using Millipore syringe filter and 10オl injected in to the HPLC system and their chromatogram were recorded for 10min. Under the chromatographic conditions as described after getting a stable baseline. Peak area was recorded for all the peaks. Calibration curve of Ritonavir was constructed by plotting the peak area of Ritonavir v/s Conc. of Ritonavir. The correlation coefficient (r2) of least square linear regression for Ritonavir was calculated.
III. Range
The range of analytical method was decided from the interval between upper and lower level of calibration curves by plotting the curve. The correlation coefficient (r2) of least square linear regression for Ritonavir was calculated.
IV. Accuracy
The accuracy of an analytical parameter is the closeness of the test results to true value. 0.75ml, 1 ml and 1.25ml from sample stock solution were transferred to three different 10 ml volumetric flasks and volume adjusted up to 10 ml with diluent as an mobile phase.
All the solutions were filtered through Millipore syringe filter and injected into the HPLC system and chromatograms were recorded under the same chromatographic conditions after getting a stable baseline. Peak area was recorded for all the peaks. From above data percentage recoveries were calculated.
V. Precision
The precision of an analytical method was studied by performing Repeatability and intermediate precision.
a) Repeatability:
100オg/ml of Ritonavir solution was filtered through syringe filter and 10オl injected into the HPLC system and its chromatogram was recorded under the same chromatographic conditions after getting a stable baseline. Peak area was recorded. The procedure was repeated for five times and the % RSD was calculated.
VI. Limit of Detection and Limit of Quantitation
a) LOD calculated by the following formulae.
LOD = 3.3(SD/S) Where, SD- Standard deviation; S- Slope of Curve.
b) LOQ calculated by the following formulae.
LOQ = 10(SD/S) Where, SD- Standard deviation; S- Slope of Curve.
VII. System Suitability
Chromatograms were studied for different parameters such as tailing factor, resolution and theoretical plates to see that whether they comply with the recommended limit or not.
4.0 RESULTS AND DISCUSSION:
4.1 Method Development
The development of the RP-HPLC method for determination of drugs has received considerable attention in recent years because of its importance in routine quality control analysis. A RP蓬PLC method was proposed as a suitable method for the estimation of Ritonavir in pharmaceutical dosage form. A good separation was achieved using a ZOROBAX Bonus-RP C-18 column (250 mm x 4.6 mm, 5 オm). The chromatographic conditions were adjusted in order to provide a good performance of the assay. The method involved a mobile phase consisting of Methanol: Acetonitrile: 0.1% Trifluoroacetic acid water (81:9:10) accomplished at 250nm. The retention time was 3.05 min at a flow-rate of 1ml/min and the injection volume was 10オl. The total run time for an assay was approximately 10 min.
4.2 Validation of the method:
The optimum ratio of mobile phase for Methanol: Acetonitrile: 0.1% Trifluoroacetic acid water was finalized that (81:9:10) respectively, at the flow rate of 1ml/min. The separation was achieved using ZOROBAX Bonus-RP C18 column at the column temperature of 30ーCア0.8ーC. Sample and standard samples were injected in the volume of 10オl via autosampler. The retention time of Ritonavir in this proposed method was found to be 3.05 minutes.
I. Specificity and Assay
Fig. 2 Chromatogram of Ritonavir
Table No. 1 Assay Result
|
Sample |
Area |
Assay (%) |
|
WS |
1088559 |
- |
|
DP |
1071861 |
98.47 |
With above optimized conditions Ritonavir was eluted at 3.05min (Fig. 2). The sample solution was assayed at 250nm & the results were found to be 98.47% w/w (Table No. 1) which complies within the range of 99.5-100.5% w/w. Hence the method was specific, peak was sharp, and with good resolution. It eluted within 5 minutes, which reduced the analysis time and cost.
II. Linearity:
Fig. 3 Overlain Chromatograms of serial dilutions of Ritonavir in optimized chromatographic conditions
The peak response is directly proportional to the concentration of drug and was found to be linear in the range of 50-150オg/ml (Fig. 3). The correlation coefficient was found to be 1 which is well within the acceptance criteria.
Table No. 2 Response of Ritonavir at various linearity levels
|
% Level |
50 |
75 |
100 |
125 |
150 |
|
Conc. (ug/ml) |
50 |
75 |
100 |
125 |
150 |
|
Area |
545424 |
817128 |
1085226 |
1359757 |
1631497 |
Fig. 4 Calibration curve of Ritonavir of RP-HPLC method
III. Range
The Linearity range was from 50-150オg/ml of the test concentration.
IV. Accuracy
Table No. 3 Accuracy for RP-HPLC Method
|
Sample ID |
Conc. (ug/ml) |
Area |
Amount Recovered (ug/ml) |
% Recovery |
Average |
STDEV |
RSD |
|
75_Rep 1 |
74.85 |
817128 |
74.9149513 |
100.09 |
99.99 |
0.133727 |
0.13 |
|
75_Rep 2 |
74.85 |
815584 |
74.7733961 |
99.90 |
|||
|
100_Rep 1 |
99.80 |
1085226 |
99.4943912 |
99.69 |
99.76 |
0.094969 |
0.10 |
|
100_Rep 2 |
99.80 |
1086688 |
99.6284285 |
99.83 |
|||
|
125_Rep 1 |
124.75 |
1359757 |
124.663614 |
99.93 |
99.90 |
0.047965 |
0.05 |
|
125_Rep 2 |
124.75 |
1358834 |
124.578992 |
99.86 |
The percentage recoveries of the results indicate that the recoveries are well within the acceptance range, therefore; method was found to be accurate.
V. Precision
a. Repeatability
Table No. 4 Results of Repeatability Study for Ritonavir
|
Injection |
1 |
2 |
3 |
4 |
5 |
AVG |
SD |
RSD |
|
Peak Area |
1085226 |
1086688 |
1088955 |
1091654 |
1090274 |
1088559 |
2611.829 |
0.24 |
The percentage RSD (<2) values obtained showed that the method developed was precise at repeatability and intermediate precision level.
VI. Detection Limit and Quantification Limit
Limit of detection (LOD) and quantification (LOQ) were estimated from the signal-to-noise ratio. LOD and LOQ values were found to be 0.68 オg/ml and 2.07オg/ml, respectively. Detection Limit and Quantification limit was calculated on basis of standard deviation of response and slope.
VII. System Suitability Testing
Table No. 5 Results of System Suitability Parameters
|
Analyte |
Retention Time (min) |
Tailing Factor (T) |
Theoretical Plates (N) |
|
Ritonavir |
3.05 |
1.07 |
9044 |
|
Required limits |
-- |
T < 2 |
N > 2000 |
Study of retention time, tailing factor and capacity factor showed that the system was suitable for the developed method.
6.0 CONCLUSION:
Thus, it can be concluded that the method developed in the present investigation was simple, sensitive, specific, linear, accurate, rapid and precise. The method exhibits simplicity in terms of short analysis time, gradient mode of elution of mobile phase, effective and clear resolution with low LOD and LOQ values. The proposed method was developed and validated as per the ICH guidelines.
7.0 ACKNOWLEDGEMENT:
We are thankful to the Principal College of Pharmacy, Solapur and Aadhar Life Sciences Pvt. Ltd., Solapur for providing required facilities for research work.
8.0 REFERENCES:
1. Indian Pharmacopoeia, vol. III, 2007, 1058.
2. United States Pharmacopoeia 30, National Formulary 25, 2007, 3143.
3. www.rxlist.com
4. Richard M, Hoetelmans W, Marjolijn Van Essenberg, Monique Profijt, Pieter L. Meenhorst, Jan W. Mulder. High-Performance Liquid Chromatographic determination of Ritonavir in human plasma, cerebrospinal fluid and saliva. Journal of Chromatography B. 1998;705(1):119-126.
5. Rebiere Herve, Mazel Bernard, Civade Corinne, Bonnet Pierre-Antoine. Determination of 19 antiretroviral agents in pharmaceuticals or suspected products with two methods using High-Performance Liquid Chromatography. Journal of chromatography B. 2007;850:376-383.
6. Yekkala RS, Ashenafi D, Marien I, Xin H, Haghedooren E, Hoogmartens J et al. Evaluation of an International Pharmacopoeia method for the analysis of Ritonavir by Liquid Chromatography. Journal of pharmaceutical and biomedical analysis 2008; 48(3):1050-4.
7. Veronica Albert, Pilar Modamio, Cecilia FL and Eduardo LM. Determination of Saquinavir and Ritonavir in human plasma by RP-HPLC and the analytical error function. Journal of Pharmaceutical and Biomedical Analysis. 2004;36(4):835-840.
8. Usami Yoshiko, Tsuyoshi OK, Naka Masahiko, Sagisaka Masafumi, Kaneda Tsuguhiro. A simple HPLC method for simultaneous determination of Lopinavir, Ritonavir and Efavirenz. Journal-Chemical and pharmaceutical bulletin. 2003;51:715-718.
9. Dias CL, Rossi RC, Donato EM, Bergold AM and Froehlich PE. LC Determination of Ritonavir, a HIV Protease Inhibitor, in Soft Gelatin Capsules. J chromatographia. 2005; 62:589-593.
10. Proust V, Toth K, Hulin A, Taburet AM, Gimenez F, Singlas E. Simultaneous High-Performance Liquid Chromatographic determination of the antiretroviral agents� Amprenavir, Nelfinavir, Ritonavir, Saquinavir, Delavirdine and Efavirenz in human plasma. Journal of chromatography B. 2000;742:453-458.
11. K.Chiranjeevi, K.P.Channabasavaraj, P.Srinivas Reddy, P.T.Nagaraju. Development and Validation of Spectrophotometric Method for Quantitative estimation of Ritonavir in Bulk and Pharmaceutical Dosage Forms. International Journal of Chemtech Research. 2011;3(1):58- 62.
12. K. Chiranjeevi, G. Vijaya Kumar, K.P. Channabasavaraj, Y. Manjunath. Third Order Derivative Spectrophotometric Estimation Of Ritonavir In Bulk And Pharmaceutical Dosage Forms. Int.J.Ph.Sci. 2011; 3(1):1017-1020.
13. Dias CL, Bergold AM, Pedro EF. UV-Derivative Spectrophotometric Determination of Ritonavir Capsules and Comparison with LC Method. Anal Letrs. 2009; 42(12):1900-1910.
14. Vaishali PN, Kishore PB. Simultaneous Estimation of Ritonavir and Lopinavir by Absorption ratio (Q-analysis) UV Spectrophotometric Method in Combined Tablet Dosage Form. Der Pharmacia Lettre. 2010;2(1):196-200.
15. Temghare GA, Shetye SS, Joshi SS. Rapid and Sensitive Method for Quantitative Determination of Lopinavir and Ritonavir in Human Plasma by Liquid Chromatography Mass Spectrometry. E-Journal of Chemistry 2009;6(1):223-230.
16. Yadav PK, Chandanam S, Rao ST, Akkamma HG, Bharadwajan RG, Deepa MS, New analytical method development and validation of acyclovir by RP-HPLC. Unique J Pharmaceutical and Biological Sciences. 2016;04(02):20-26.
17. Sulebhavikar AV, Pawar UD, Mangoankar KV, Prabhunavelkar ND. HPTLC Method for Simultaneous Determination of Lopinavir and Ritonavir in Capsule Dosage Form. E-Journal of Chemistry 2008;5(4):706-712.
Received on 14.10.2020�������� �� Modified on 11.11.2020
Accepted on 07.12.2020�������� � ゥ RJPT All right reserved
Research J. Pharm. and Tech 2021; 14(10):5457-5460.
DOI: 10.52711/0974-360X.2021.00951